Midas Dl 371 User Manual Psf

User manual 8.5 FILE SERVER EDITION - V 85.5 Specifications 9.0 EMAIL SERVER EDITION - REVISED 10-2009 User manual 9.0 FILE SERVER - REV 90.1 Specifications 9.0 FILE SERVER - V 90.8 User manual Linux E-mail Server User manual Avid Technology AirSpeed 5000 Setup guide. Product specifications AS3000.

Download MIDAS manuals for offline viewing

As well as the extensive online help documentation (which is also included within each MIDAS room booking system itself via the inbuilt [Help] links),
we are pleased to be able to additionally offer complete manuals in PDF format for you to download, print, and view offline.

ManualDescriptionDateSizeView/Download
MIDAS Manual v4.22Complete 87-page User and Administrative Guide to MIDAS v4.2217 July 20191.22MB
API Documentation v2.19Complete API documentation12 August 2017923KB
Digital Signage Documentation v1.06Complete documentation for our optional Digital Signage addon5 May 2014707KB
LDAP Integration Documentation v1.03Guide to integrating MIDAS with your Active Directory,
providing seamless Single Sign-On ability
31 July 2017754KB
One moment please...

EMD Chemicals Inc. - Midas III Stainer
by EMD Chemicals Inc.

DESCRIPTION
The Midas III Stainer is an automated slide stainer designed for use in hematology and microbiology laboratories.

The Midas III Stainer is an automated slide stainer designed for use in hematology and microbiology laboratories.It features compact design and microprocessor control for flexibility in staining applications.Up to 30 slides or coverslips can be stained uniformly in minutes. The Midas®III Stainer has nine user definable programs each containing up to 29 possible processing steps.For added convenience, it performs an automatic electronic self-test when turned
on and a mechanical self-test as required by the operator. An optional recirculating Water Bath is available for use with the Midas®III Stainer and should be used in hematology unless a dedicated deionized water line is available.A water bath can be used in microbiology. (Tap Water Desirable For Microbiology)

FORUMSView All (10)
Ask a New Question

1

Reply
-Marc Kidney Tech
2 months ago
2 months agoWont go lower
Device rest position proper when started up down and left right.But wont go down on program. New up and down motor was installed did work once then issue startedReply
-ShaunBIO
2 years ago
2 years ago Schematics
Do you have the schematics to take apart and replce the keyboard?Reply

0

Replies
-owcn
3 years ago
3 years agoowners manual Midas III hematology stainer

Does anyone have an electronic version of the owner's manual for the Midas III slide stainer? Can't find one online. Thanks!

Reply
DOCUMENTS / MANUALSView All


FEATURES

SYSTEM FUNCTION Specimen slides are loaded into the removable slide carrier, and the carrier is mounted onto the carrier arm of the tower.Stains and fixatives are added to staining vessels and inserted into the Stations on the stainer. During operation, the tower holding the slide carrier moves from Station to Station according to a programmed sequence.The carrier arm dips the slides by moving up and down automatically every three seconds in the staining vessel.This Dipping Mode can be turned on or off for Stations 1 through 6 by the operator.If Station 5 has been programmed, the auto-rinse feature turns the water on and then off when the staining cycle is complete.If Station 6 has been programmed, the slides are dried with warm air. (Temperature adjustable from room temp to 65ºC) When the program is complete, the tower moves back to the far left, its Home Position.The Midas®III Stainer gives an audible beep six times to let the operator know that the process is complete. FUNCTIONAL CHARACTERISTICS

  • Staining Control Microprocessor * Dipping mechanism ON/OFF selectable at all Stations * Available Programs 9 user programmable, 29 steps in each; 1 test program * Staining Vessel Capacity 300ml * Available Stations 6 in total, 4 staining/reagent, 1 running water rinse, 1 warm forced-air dry or 6 staining/reagent * Stations * Available Program * Application Time 1 second to 99 minutes, 59 seconds per program step with 29 steps per program * Slide Carrier Capability 20 slides, 3 x 1 inch or optional 30 slide carrier * Coverslip Carrier Capability 20 coverslips, 22 x 22mm * Rinse Station Flow Rate Application range - 500-2000ml/min * Available Options Recirculating Water Bath, Product No. 64000WB-94. * Vessels and Lids - Neutral White - package of 8, * Product No. 64054W-95 * Vessels and Lids - Black - package of 8, * Product No. 64054B-95 * Coverslip Carrier, Product No. 64053-94. * Optional 30 Slide Carrier, Product No. 64052E-94

Additional Specifications

ELECTRICAL CHARACTERISTICS



Power Input: Voltage: AC, 115 volts or 230 volts (external switch
selectable), 50 or 60 Hz


Current: 5 Amps max for 120 volts input, 5 Amps max for 230 volts input


Watts: 600 watts max at 120 volts


Fuse Rating: 5 Amp x 250 volts for both 110 volts and 230 volts units (5A Amp Max Fuse 5A 250 VAC +/-Slo-Blo)


Heating station: Heater:375 watts, AC driven


Blower:less than 15 watts, AC driven (CB1Fan)


Internal Circuit Protection:4 Amp circuit breaker (CB2Heater)


ManualRinse station: Water valve:24 volts DC, 1 Amp max, proportional valve (CB1Fan, GFI)


Power Output:GFI protected power outlet for external water bath pump, 120 volt AC output, auto-transformer output.


Internal Circuit Protection: 1 Amp circuit breaker (CB1, GFI)


Display: 5 volts DC, max 0.5 amp, vacuum fluorescent display, 2 lines x 16 characters per line presentation


Keyboard: Membrane Keyboard, 20 keys


Midas Dl 371 User Manual Pdf Download

PHYSICAL CHARACTERISTICS
  • Length: 21' (53cm)
  • Width: 18' (45cm)
  • Height: 15' (38cm)
  • Weight: 28 lbs (13kg)


    ENVIRONMENTAL CONDITIONS
    Midas®III Stainer is a laboratory instrument and should be used in a medical/pathology laboratory with the following conditions:
  • Ambient Temperature: 15ºC to 40ºC
  • Relative Humidity: 20-85% non-condensing
  • Midas Dl 371 User Manual Pdf Online

  • Ambient Pressure: 28'-32' (70-80mm) of Hg
    INSTALLATION CATEGORY
    In reference to Installation category as defined in UL 3101-1 standard, Midas®III Stainer is defined as a Category II system
    POLLUTION DEGREE
    In reference to Pollution Degree as defined in UL 3101-1 standard, Midas®III Stainer is categorized as a Pollution Degree 2 system

    MECHANICAL CHARACTERISTICS
    Enclosure:
    Two parts, Both Top and Bottom Enclosures are made of Kydex®thermoplastic sheet.

    Instruction Manual

    Reagent Containers: All reagent containers (including Drying and Rinse Station containers) are made of HDPE. (High Density Polyethylene)
    Rear Panel:

    User Manual

    Aluminum